FOUNDATION Fieldbus Communication Introduction
Table 8 - Maximum Static/Overrange and Proof Pressure for DP and Biplanar
Transmitters
Transmitter Configuration
18
Maximum Static
19
and
Overrange Pressure
20
Maximum Proof
Pressure
21
Standard (B7 steel) with Span Codes B to
E, or with Option -B2 (17-4 PH ss), -D3,
-D7, -P3, or -P7
25 MPa (3,626 psi) 100 MPa (14,500 psi)
Standard with Span Code F
22
40 MPa (5,800 psi) 100 MPa (14,500 psi)
Option -B3 (B7M), -P4, or -P8 20 MPa (2,900 psi) 70 MPa (11,150 psi)
With Option -D1 16 MPa (2,320 psi) 64 MPa (9,280 psi)
Option -B1 (316 ss), -D5, -P2, or -P6 15 MPa (2,175 psi) 60 MPa (8,700 psi)
With Option -D2, -D4, -D6, or -D8
23
10 MPa (1,500 psi) 40 MPa (6,000 psi)
Option -D9 (17-4 PH ss) or -Y
24
40 MPa (5,800 psi) 100 MPa (14,500 psi)
With Structure Code 78 or 79 (PVDF
insert)
25
2.1 MPa (300 psi) 8.4 MPa (1,200 psi)
Elevated Zero and Suppressed Zero
For applications requiring an elevated or suppressed zero, do not exceed the
maximum span and the upper and lower range limits of the transmitter.
Sensor Fill Fluid
Silicone fluid dodecamethylpentasiloxane
3M
Fluorinert
Electronic Liquid FC-43 perfluorotributylamine
NEOBEE
®
M-20 propylene glycol di(octanoate/decanoate)
Minimum Allowable Absolute Pressure vs Process Temperature
With silicone fill fluid: up to 121°C (250°F) at full vacuum
26
With inert fill fluid: refer to the graph
26 27
16 MI 020-612
18. Refer to the model code for option descriptions, and for applications and restrictions related to the items listed in the table.
19. Static pressure is relevant only for differential pressure transmitters.
20. Either side can be at higher pressure during overrange.
21. Meets ANSI/ISA Standard S82.03-1988.
22. Available for gauge pressure transmitters only.
23. Limited to operating temperatures ranging from -10 to +80°C (14 to 176°F).
24. Differential pressure transmitters only.
25. With PVDF insert, temperature limits are -7 and +82°C (20 and 180°F).
26. For direct connect IGP50S transmitters with Span Code D, the minimum allowable pressure is 0 psig.
27. For biplanar IAP50S transmitters, up to 50°C (120°F) at full vacuum.