Voltcraft EL-1400 handleiding
Handleiding
Je bekijkt pagina 15 van 38

1
Press Enter key to save current setting.
3. Press ON/OFF key to output the setting current.
In constant current mode, the user can also set the following parameters.
Operation steps:
1. Press CONF to enter PARAM.(PARAMETER SETUP)
2. Set I_LIM (maximum working current)
3. Set V_RANGE, I_RANGE (voltage/current range)
4. Set SR_RISE(the rising slope of current)
5. Set SR_FALL(the falling slope of current)
6. Set OVP, OCP, OPP
4.3.2 Constant Voltage Mode (CV)
When the user selects the CV function key, the load enters constant voltage
mode. In constant voltage operation mode, the electronic load will draw
sufficient current to maintain the input voltage at the set value. The
relationship between voltage and current is shown in the figure below.
Load
voltage
Load current
Setting
voltage
Figure 4.3.2-1 CV mode: The Voltage-Current Relationship Diagram