The probes have no user- or Tektronix-serviceable parts.
Warning: Do not float the
TPP0100, TPP0200 probes on any oscilloscope.
Connecting the probe to the oscilloscope
Connect the probe as shown in the illustrations below
.
Compensating the probe
Due to variations in oscilloscope input characteristics, the low-frequency compensation of the probe may need adjustment after moving the
probe from one oscilloscope channel to another
.
If a 1 kHz calibrated square wave displayed at 1 ms/division shows significant differences between the leading and trailing edges, perform
the following steps to optimize low-frequency compensation:
1. Connect the probe to the oscilloscope channel that you plan to use for your measurements.
2. Connect the probe to the probe compensation output terminals on the oscilloscope front panel.
Warning: T
o avoid electric shock, only connect to the Probe Comp signal on the oscilloscope when making this adjustment.
3. Push Autoset or otherwise adjust your oscilloscope to display a stable waveform.
4. Adjust the trimmer in the probe until you see a perfectly flat-top square wave on the display
. (See illustration.)
Warning: To avoid electric shock, only use the insulated adjustment tool when making compensation adjustments.
Operating requirements
TBS1000C Series Oscilloscopes User Manual 28