Operating requirements
Environment requirements
Characteristic Description
Operating and Non-Operating
temperature
Operating: 0°C to +50°C
Non-Operating:
-30°C to +7
0
°C, with 5°C/minute maximum gradient.
Operating and Non-Operating
humidity
Operating:
5% to 90% relative humidity (% RH) at up to +30°C
5% to 60% RH above +30°C up to +50°C
non-condensing
Non- Operating:
5% to 90% relative humidity (% RH) at up to +30°C
5% to 60% RH above +30°C up to +60°C
non-condensing
Operating and Non-Operating
altitude
Operating: Up to 3,000 meters (9,842 feet)
Non-Operating:
Up to 12,000 meters (39,370 feet)
Power requirements
Characteristic Description
Power source voltage 100 to 240 V
AC
RMS ±10%, Single Phase
Power source frequency 50/60 Hz over entire source voltage range
400 Hz (360 Hz to 440 Hz) for 1
15 V
AC
(100 V
AC
- 132 V
AC
) RMS source voltage range
Power consumption All models: 80 W maximum
CAUTION: A ground connection through the power cord grounding conductor is essential for safe operation.
TPP0100, TPP0200 series 10X passive probes information
The TPP0100, TPP0200 Series 10X Passive Probes are high impedance, passive probes with 10X attenuation. They are designed for use
with TBS1000C oscilloscopes. These oscilloscopes have
14 pF
of input capacitance.
The compensation range of these probes is 8 to 18 pF.