XPT800 Sound Level Meter and Spectrum Analyzer
V1.0
Page 91 of 124
13.2 Firmware update
Updating the firmware requires a PC with a USB port and the file containing the new firmware.
Firmware files are of the type:
file_name.hex
1
Turn on and connect the XPT800 Sound Level Meter to the
USB port of your PC using USB-C cable.
2
In SYS-FUNC/USB/PROCEDURES press FILE WRITE to activate
USB write (don’t exit the pop-up!)
3
As soon as access to the memory unit is available, copy the
"new_firmware.hex" file to the folder > Firmware.
4
On the FILE WRITE pop-up press EXIT to deactivate USB write
5
Switch the instrument off and on again
6
In ICON MENU, press the SYS-FUNC icon to access the sys-
tem and measurement configuration menus.
7
Access the FW UPGRADE and PROCEDURES pages in suc-
cession.
8
Tap on LOAD NEW.
The device automatically reboots until the splash screen is displayed; the status LED starts blinking
green. The update procedure can take between 10 and 20 minutes.
Verify the correct update of the new version as described in chapter 13.1 Firmwareon page 90.
Info
In the case that the update cannot be carried out, e.g. due to a battery voltage below 3.8V, a
running measurement, or a missing correct .hex file, a pop-up alerts the user with a message.
Notice!
In the event that the firmware update fails, the status LED turns red, indicating a possi-
ble error in the upload procedure...
The instrument has a recovery firmware pre-loaded in a non-editable memory area; to
restore the firmware, please refer to section 13.3 Instrument recovery via recovery
firmwareon page 93.