Senseca PRO 241-2 handleiding
Handleiding
Je bekijkt pagina 19 van 27

PRO 241-2 / PRO 245-2 – Absolute manometers
V10
Page 19 of 27
9 Logging (PRO 245-2 only)
The logging function stores in the memory of the instrument the detected measurements, along
with the date and time of each sample. Data recording can be automatic, according to the set
interval, or single (manual).
Attention!
• To use the logging function, the USB mode menu item must be set to COM interface.
• The Logging function records only the quantities enabled in the measurement screens.
Before starting logging, make sure that all quantities of interest are enabled.
9.1 Automatic logging
The automatic logging data are stored in CSV format.
To start automatic logging, select the Start recording item from the Recording menu.
By selecting "Start recording," the instrument automatically exits the menu. It is not possible
to access the menu during logging.
The logging activation is highlighted by the rec symbol in the status bar of the display. During
logging, the status LED briefly flashes green every 5 seconds.
With logging active, in the measurement screens the left function key assumes the function of
Stop logging.
During logging, the instrument auto-off is disabled.
9.2 Single recording
The single recordings data are stored in a single file in JSON format (history.json). The new re-
cordings are added to existing recordings, which are not deleted.
Three modes are available for single recording, which can be selected in the Recording →
Mode menu:
• Standard
• Multiselect
• On hold
Standard mode:
To make a single recording, select the Single recording item of the Recording menu, then se-
lect with the down/up arrow keys one of the available labels (Location xx) and press OK to
confirm and save the measurement data.
During data saving, the rec symbol lights up in the status bar of the display.
The selected label is marked with the check mark. The label associated with the recording is
useful for distinguishing the various recordings in the data file, which is unique for all single re-
cordings. The label may, for example, represent the location where the recording takes place;
recordings made in different locations may be associated with different labels. There are 20 dis-
tinct labels available (Location 1...20).