7 Measurement basics
The instrument internal pressure sensor detects the absolute pressure applied to the input.
Caution!
• Do not exceed the maximum overpressure specified, the sensor may be damaged.
• The sensor is suitable for measurement in air and non-aggressive dry gases: do not use for
measurement in corrosive gases.
• Avoid performing measurements in the presence of high-frequency sources, microwaves
or strong magnetic fields.
• Connect the hose to the instrument without forcing or bending the pressure port.
8 Measurement adjustment
The measurement can be adjusted by setting the following parameters, available in the Pres-
sure sensor → Correction:
• Zero point: adds an offset (P
offs
) to the measurement.
Displayed value = Measured value + P
offs
• Gradient: applies a percentage correction (C
slope
%) to the measurement. Zero indicates
no correction.
Displayed value = (Measured value + P
offs
) * (1 + C
slope
% / 100)
If a measurement correction is applied by setting a value other than zero for the “Zero point"
and/or "Gradient” parameters, the cor symbol appears on the display.