7 Measurement basics
The instrument internal pressure sensor detects the absolute pressure applied to the input.
Caution!
Do not exceed the maximum overpressure specified, the sensor may be damaged.
The sensor is suitable for measurement in air and non-aggressive dry gases: do not use for
measurement in corrosive gases.
Avoid performing measurements in the presence of high-frequency sources, microwaves
or strong magnetic fields.
Connect the hose to the instrument without forcing or bending the pressure port.
8 Measurement adjustment
The measurement can be adjusted by setting the following parameters, available in the Pres-
sure sensor Correction:
Zero point: adds an offset (P
offs
) to the measurement.
Displayed value = Measured value + P
offs
Gradient: applies a percentage correction (C
slope
%) to the measurement. Zero indicates
no correction.
Displayed value = (Measured value + P
offs
) * (1 + C
slope
% / 100)
If a measurement correction is applied by setting a value other than zero for the Zero point"
and/or "Gradient parameters, the cor symbol appears on the display.