Senseca PRO 115 handleiding
Handleiding
Je bekijkt pagina 17 van 25

PRO 111 / PRO 115 – Pt100 thermometers
Page 17 of 25
7.2 Warning on USB port isolation
The instrument USB port is not galvanically isolated, and its connection, either to the PC or to
an external power supply that is not isolated from ground, may affect the measurement when
using a probe that is in contact with the measured medium. Under such conditions, always
check the measurement with and without USB connected: if a difference in the detected value
is noticed, disconnect USB, or use an external suitable USB isolator (for example, DFRobot
FIT0860) to detect reliable measurements.
Bekijk gratis de handleiding van Senseca PRO 115, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.