7 Measurement basics
The sensor is located at the end of the probe.
Immersion or penetration measurement: immerse the probe stem for at least 60 mm; when
measuring in liquids, stir the liquid if possible.
Attention!
The measurement may be inaccurate if the probe stem is not sufficiently immersed, due to
heat loss from the metal stem.
Measurement in air/gases: immerse the stem as much as possible in the fluid to be measured.
The response time is shorter in the presence of flow; in the absence of flow, you can speed up
the measurement by moving the probe, if possible.
Attention!
The probe stem must be dry, otherwise a temperature lower than the actual temperature
will be detected.
Measurement of solid materials by contact: the measuring surface should be flat and smooth;
the probe should be perpendicular to the measuring plane. Probes with tip specially designed
for contact measurements must be used. The ambient temperature and heat dissipation of the
probe metal stem can affect the accuracy of the measurement.
Attention!
For a more accurate and faster contact measurement, interpose thermally conductive paste
between the measuring surface and the probe tip.
Measurement on non-metallic surfaces takes longer because of the poor thermal
conductivity.
7.1 General warnings on probes usage
Wait for thermal equilibrium between the sensitive part of the probe and the area to be meas-
ured before taking the measurement.
Thoroughly clean the probe after use.
Danger!
If the probe has a metal stem or other metal parts, be careful not to come into contact with live
parts.
Caution!
Do not expose the probe to corrosive gases or liquids!
Do not expose the probe to temperatures exceeding the operating limit specified for the
probe, the measuring sensor may be damaged.
Avoid performing measurements in the presence of high-frequency sources, microwaves
or strong magnetic fields.
Connect the probe to the instrument without forcing or bending the connectors or
contacts!
Do not deform or drop the probe!
Bending within moderate radius is only allowed for mineral insulated probes.