ECO 230 barometer | altimeter
Measurement Basics
B-H88.0.01.DK2-2.1 Page 12 of 20
5 Measurement Basics
5.1 General information about absolute pressure measurement
The device measures absolute pressure. However, this should not be confused with the "sea
level air pressure" indicated by weather stations. The altitude-based air pressure decrease is
calculated for these pressure specifications.
The device is capable of correcting this air pressure altitude correction.
To do this, set the parameter value SEA.L in the Configuration menu to YES and enter the
current altitude above sea level in the parameter value ALT,.
The nautical norm correction must be deactivated for vacuum measurements.
5.2 Altimeter
If meters or feet are selected as the display unit in the Configuration menu, the device is in
altimeter mode. The unit calculates the current altitude from the current air pressure: At
higher altitudes, the air pressure is lower. It is important to note that not only changes in
altitude but also the weather influence the measurement. To correct this weather influence,
the displayed altitude can be corrected via the keys.
Altimeter display: In the Altimeter operating mode, you will be prompted to enter the
current altitude after pressing the Function key for 4 s.
Note: The first time you are prompted after switching on the device, the value is set
to the alti value in the Configuration menu. (default value for starting point)
The current value is retained for all further calls.
Note on min/max display: in the secondary display, the sum of the altitude meters covered
is also shown for min: dESC - descent, for max: ASC - ascent
5.3 Special functions
5.3.1 NVLL Tare function
If a pressure unit is selected as the display unit in the Configuration menu, the device is in
barometer mode.
After it is switched on, the device starts up in standard measuring mode, the special function
“tare” is started by pressing and holding the Function key for 4 s..
Barometer display: The display can be zeroed by pressing the Function key for 4 s. If the
tare function is activated, NVLL blinks in the lower dis-play. The tare
function can be reset by pressing the Function key again for 4 s.