ECO 230 barometer | altimeter
Measurement Basics
B-H88.0.01.DK2-2.1 Page 13 of 20
Note
The tare function is independent of the zero point correction accessible via the adjustment
menu.
5.4 Use of the trend indicator
Note
For the trend display, it is necessary that the device has collected a sufficiently large
number of measured values. This is the case for the time bases 1h and 3h only after this
time has elapsed.
The value is therefore not shown in the display before this time has elapsed.
5.4.1 Meteorology: weather forecasting
Observation of variable weather conditions by assessing the rate of change of air pressure.
To do this, set the device to:
UN,T
HPa
(international standard)
RATE
SLO
(power saving during continuous operation)
SEA.L
YES
ALT,
altitude of the location above sea level
TEND
1_H or 3_H
1 hour (1 hPa / bar level)
or 3 hours (3 hPa / bar level)
LCD.2
°(.TE
POFF
OFF
continuous operation
This results in the following display, for example:
display unit hPa
current pressure at sea level
trend: + 0.4 hPa per hour
□■■□ stable (trend lower 1 hPa)