ECO 230 barometer | altimeter
Operation
B-H88.0.01.DK2-2.1 Page 11 of 20
Parameter
Values
Meaning
IN,T
Factory settings
NO
Use current configuration
YES
Reset device to factory settings.
After confirming with the function-button, the
display shows: IN,T DONE
4.2 Open the adjustment menu
The sensor adjustment can be adjusted with the zero point correction and the gradient
correction. If an adjustment is made, you change the pre-adjusted factory settings.
This is signaled with the display text PR.OF or PR.SL when switching on.
1 Switch the device off.
2 Hold the down button and press the On/Off button to switch on the device and open the
Adjustment menu.
3 The display shows the first parameter. Release the down button.
Parameter
Values
Meaning
PR.OF
Zero point correction
0.0
No zero point correction
-5.0 5.0
Zero point correction [in selected display unit]
(e.g. ± 5.0 hPa)
PR.SL
Gradient correction
0.00
No gradient correction of the temperature
-5.00 5.00
Gradient correction in %
Formula:
Zero point correction: Displayed value = measured value - PR.OF
Gradient correction: Display = (measured value - PR.OF) * (1 + PR.SL / 100)