33 
select the desired language using the left/right arrow key. 
7  How do I view the information and model number of the plug-in 
modules? 
Press Utility  Test; select the desired module using the left/right arrow key 
and press View to view the information and model number of the module 
currently selected. 
8  The corresponding indicator on the front panel does not go on when 
the module has been inserted. 
1)  Press Utility  Test; select the desired module using the left/right arrow 
key and press Re-test.   
2)  If the problem persists, press View. If the information of the module 
cannot be viewed, please refer to the Problem 9. 
3)  If the information of the module can be viewed, the module indicator is 
possibly damaged. Please contact RIGOL for maintenance information. 
9  I cannot view the information of the module or use the module when 
the module has been inserted. 
1)  Turn the main frame off and remove the module from the main frame; 
insert the module into the mainframe again firmly. 
2)  Restart the main frame, wait for about 1 minute and try to view the module 
information again. 
3)  If the problem persists, please contact RIGOL. 
10  I cannot make configurations or operations related to the DMM 
module when it has been inserted. 
1)  Make sure that Utility  System  DMM is set “ON”.   
2)  Press Utility  Test; select the DMM module using the left/right arrow key 
and press View to view whether the information of the DMM module is 
normal. If the information is not displayed or is not displayed correctly, it is 
possibly that the DMM module is working abnormally due to high current or 
power load failure. At this point, please turn the main frame off and restart 
it a few minutes later.   
3)  If the problem persists, please contact RIGOL. 
11  The GPIB interface cannot work normally. 
1)  Check whether the cable works normally and the connection is correct and 
reliable. 
2)  Make sure that the GPIB addresses of M300 and the PC are identical. 
3)  If the problem persists, please contact RIGOL. 
12  The USB Device interface cannot work normally. 
1)  Check whether the cable works normally and the connection is correct and 
Bekijk gratis de handleiding van Rigol M302, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.