RIGOL
M300 Quick Guide
34
reliable.
2) Check whether the USB Test and Measurement Device (IVI)drive
program has been correctly installed on your PC.
3) If the problem persists, please contact RIGOL.
13 The RS232 interface cannot work normally.
1) Make sure that the RS232 interface parameters (such as the baud rate)
settings of M300 and the PC are identical.
2) Check whether the RS232 cable is a cross cable.
3) If the problem persists, please contact RIGOL.
14 The LAN interface cannot work normally.
1) Check whether the cable works normally and the connection is correct and
reliable.
2) Check the IP configuration mode:
If DHCP is enabled, please make sure that the current network
supports DHCP configuration mode and can distribute network
parameters (such as the IP address) to M300 automatically.
If DHCP is disabled and Auto IP is enabled, please check whether the
IP addresses automatically acquired by M300 and the network
parameters of the PC are in the same network segment.
If only Manual IP is enabled, please make sure that the network
parameter settings of M300 and the PC are in the same network
segment.
3) If the problem persists, please contact RIGOL.