RaySafe X2 SURVEY
Manual trig mode
Time is calculated as the time from when you press record until you press stop.
Dose is calculated from all recorded data.
Dose rate peak (ˆ) is the highest dose rate recorded during measurement. A waveform is dis-
played if the peak dose rate exceeds 1
µ
Sv/h or 1
µ
Gy/h (0.1 mR/h).
Mean energy is calculated from all recorded data.
SURVEY SENSOR: SPECIFICATIONS
Dimensions: 14 × 66 × 192 mm (0.5 × 2.6 × 7.6 in)
Weight: 140 g (4.9 oz)
Storage temperature: -25 +70
°
C (-13 +158
°
F)
Storage humidity: Non-condensing
Operating temperature: 15 35
°
C (59 95
°
F)
Operating atmospheric
pressure:
70 110 kPa (3000 m above sea level)
Operating humidity: < 80 % relative humidity, non-condensing
Reference point: Center of entrance window, at a depth indicated by lines at
the sides of the sensor
Direction of incident radiation: Orthogonal to the entrance window
Minimum field of uniform
radiation:
Size of entrance window; 67 × 73 mm (2.6 × 2.9 in)
Angular deviation, dose: < 1% within ± 10
°
Backscatter: Back side of sensor protected
Sound: Tick frequency proportional to the measured dose rate
Angular deviation, dose:
39