RaySafe X2 – SURVEY
SETTING: TRIG MODE
The trig mode setting affects both how a measurement is triggered and which parameters are dis-
played.
Auto
When in auto trig mode, start of measurement is triggered by radiation, with trig level (N80) 20
µ
Sv/h or 10
µ
Gy/h (1.2 mR/h). The sensor is in this mode useful as a sensitive dose rate meter,
placed in the primary X-ray field.
Parameter displayed are:
accumulated dose,
time,
instantaneous dose rate (final reading: average dose rate),
mean energy.
Manual
In manual trig mode, you can start and stop the measurement from the base unit. This mode is
suitable for measuring leakage or scattered radiation from X-ray equipment and γ-emitting iso-
topes.
Parameters displayed are:
accumulated dose,
time,
instantaneous dose rate (final reading: peak dose rate),
mean energy.
SETTING: UNITS
Select displayed unit for dose and dose rate.
Air kerma, K
air
, is measured in Gy or R, where 1 Gy = 114.1 R.
Ambient dose equivalent, H
∗
(10), is measured in Sv.
Typical response:
40