RaySafe Pro-Slit Camera User Manual – Test Procedure and Evaluation
11. To calculate focal spot dimensions, divide the measured width values by the calculated enlargement factor.
If the measured width b
w
=2.9 mm (1st exposure), the focal spot width f
w
=2.9/2.2=1.32 mm
If the measured width b
l
=4.1 mm (2nd exposure), the focal spot length f
l
=4.1/2.2=1.86 mm
12. Find the nominal focal spot value for your X-ray tube and compare your measured result with the
corresponding limiting values given in Table 2 (EN 60336) on the next page.
If the nominal value is 1.0, the calculated focal spot width 1.32 ≤ 1.40, and the calculated focal spot length
1.86 ≤ 2.00. This means performance is confirmed to be ok.
Note! For some applications, the nominal focal spot value declaration consists of a pair of numbers,
e.g. 1.0 x 1.6. Then use the smaller nominal spot value, the width, when checking against Table 2.