Problem Why it Happened What to Do
I’m having difculty adjusting
the heated humidier
setting or the heated tube
temperature setting.
The blower is not turned
on, or the humidier or
heated tube is not fully
connected.
The humidier setting and tube temperature settings can only
be adjusted from the Therapy ON display screen on the therapy
device. Conrm that the blower is turned on, and that the settings
are visible on the right side of the screen, then adjust to desired
comfort. If the blower is on but the humidier settings are not
displayed on the Therapy ON screen, then unplug the device.
Conrm that the humidier and/or heated tube electrical contacts
are not obstructed or damaged. Then reconnect the humidier and/
or heated tube, and reconnect the device’s power supply. Turn the
blower on; if the settings are still not visible, contact your provider
for assistance.
The water in the water
chamber runs out before
morning.
Water chamber was not
full at start of session.
Mask leak is excessively
high. The ambient
conditions are very dry/
cool.
Under most conditions, a full water chamber should last for a typical
sleep session. However, many factors impact water consumption,
including: the ambient temperature and humidity in your bedroom,
your humidier or heated tube settings, the level of mask leak, and
the duration of your sleep session.
First, make sure that the water chamber is lled to the maximum ll
line at the start of your sleep session. Check that your mask is tted
properly, and adjust as needed to reduce mask leak to normal levels.
You may use the Check Mask Fit function to evaluate your mask
t. Also, conrm that the device, humidier, humidier seals and
tube are connected properly and not leaking. You may also choose
to lower your humidier and/or heated tube settings or change the
humidication mode from Fixed to Adaptive humidication mode to
increase the time that your humidier water will last.
I hear a leak or whistling
sound coming from my
therapy device or humidier
(not related to mask leak).
The therapy device air
inlet may be obstructed.
The humidier or tube is
not fully connected. The
humidier seals are not
fully seated or are missing.
Check therapy device air inlet is not obstructed, and lters are clean
and properly inserted. Conrm that the device, humidier, and tube
are connected properly and not leaking. Conrm that the humidier
lid seal and dry box seal are present and properly seated; if needed,
gently press around the perimeter of the seals to reseat them.
I accidentally spilled water
into my humidier basin.
The water chamber has
been lled beyond the
maximum ll line.
A small amount of water spilled in the basin of the humidier will
not harm your device. A small spill in the humidier will evaporate
under normal humidier use. However, too much water in the
humidier basin could spill over the humidier lid hinge and might
damage your furniture.
Disconnect power from the device. Remove the water chamber,
pour out any excess water until the water level is at or below the
maximum ll line and set the chamber aside. Separate the humidier
from the therapy device, and pour out the spilled water. Once
the heater plate has cooled, wipe the inside of the humidier with
a paper towel or soft cloth. If needed, dry the underside of the
humidier and conrm that your table top is dry. Reconnect the
humidier and power supply, and reinstall the water chamber.
Note: For information on troubleshooting your therapy device, see the manual included with your therapy device.