Problem Why it Happened What to Do
I’m having difculty adjusting
the heated humidier
setting or the heated tube
temperature setting.
The blower is not turned
on, or the humidier or
heated tube is not fully
connected.
The humidier setting and tube temperature settings can only
be adjusted from the Therapy ON display screen on the therapy
device. Conrm that the blower is turned on, and that the settings
are visible on the right side of the screen, then adjust to desired
comfort. If the blower is on but the humidier settings are not
displayed on the Therapy ON screen, then unplug the device.
Conrm that the humidier and/or heated tube electrical contacts
are not obstructed or damaged. Then reconnect the humidier and/
or heated tube, and reconnect the device’s power supply. Turn the
blower on; if the settings are still not visible, contact your provider
for assistance.
The water in the water
chamber runs out before
morning.
Water chamber was not
full at start of session.
Mask leak is excessively
high. The ambient
conditions are very dry/
cool.
Under most conditions, a full water chamber should last for a typical
sleep session. However, many factors impact water consumption,
including: the ambient temperature and humidity in your bedroom,
your humidier or heated tube settings, the level of mask leak, and
the duration of your sleep session.
First, make sure that the water chamber is lled to the maximum ll
line at the start of your sleep session. Check that your mask is tted
properly, and adjust as needed to reduce mask leak to normal levels.
You may use the Check Mask Fit function to evaluate your mask
t. Also, conrm that the device, humidier, humidier seals and
tube are connected properly and not leaking. You may also choose
to lower your humidier and/or heated tube settings or change the
humidication mode from Fixed to Adaptive humidication mode to
increase the time that your humidier water will last.
I hear a leak or whistling
sound coming from my
therapy device or humidier
(not related to mask leak).
The therapy device air
inlet may be obstructed.
The humidier or tube is
not fully connected. The
humidier seals are not
fully seated or are missing.
Check therapy device air inlet is not obstructed, and lters are clean
and properly inserted. Conrm that the device, humidier, and tube
are connected properly and not leaking. Conrm that the humidier
lid seal and dry box seal are present and properly seated; if needed,
gently press around the perimeter of the seals to reseat them.
I accidentally spilled water
into my humidier basin.
The water chamber has
been lled beyond the
maximum ll line.
A small amount of water spilled in the basin of the humidier will
not harm your device. A small spill in the humidier will evaporate
under normal humidier use. However, too much water in the
humidier basin could spill over the humidier lid hinge and might
damage your furniture.
Disconnect power from the device. Remove the water chamber,
pour out any excess water until the water level is at or below the
maximum ll line and set the chamber aside. Separate the humidier
from the therapy device, and pour out the spilled water. Once
the heater plate has cooled, wipe the inside of the humidier with
a paper towel or soft cloth. If needed, dry the underside of the
humidier and conrm that your table top is dry. Reconnect the
humidier and power supply, and reinstall the water chamber.
Note: For information on troubleshooting your therapy device, see the manual included with your therapy device.