Philips DreamStation handleiding

16 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 14 van 16
12
DreamStationHeatedHumidierSpecications
Environmental
Operating Temperature: 5° to 35° C (41° to 95° F)
Storage Temperature: -20° to 60° C (-4° to 140° F)
Relative Humidity (operating & storage): 15 to 95% (non-condensing)
Atmospheric Pressure: 77 to 101 kPa (0 - 2286 m / 0 - 7500 ft)
Physical
Dimensions (humidier with therapy device): 29.7 x 19.3 x 8.4 cm (11.7” L x 7.6” W x 3.3” H)
Heightdoesnotincludehumidierairoutletport.
Weight (empty humidier with therapy device and power supply): Approximately 1.98 kg (4.37 lbs.)
ServiceLife
The expected service life of the DreamStation Heated Humidier is 5 years.
Maximum Recommended Pressure
25 cm H
2
O
Water Capacity
325 ml (11 oz.) at recommended water level
Standards Compliance This device is designed to conform to the following standards:
IEC 60601-1 General Requirements for Basic Safety and Essential Performance of Medical Electrical Equipment
ISO 8185 General Requirements for Humidication Systems
Electrical (When the heated humidier is used with a Philips Respironics therapy device)
AC Power Consumption (with 80W power supply): 100 – 240 VAC, 50/60 Hz, 2.0-1.0 A
DC Power Consumption: 12 VDC, 6.67 A
Type of Protection Against Electric Shock: Class II Equipment
Degree of Protection Against Electric Shock: Type BF Applied Part
Degree of Protection against Ingress of Water: Drip Proof, IP22
Mode of Operation: Continuous
Electromagnetic Compatibility: The device meets the requirements of EN 60601-1-2.
Heater Plate
Max Temperature: 65° C (149° F)
PressureDropwithHumidier
Max: 0.6 cm H
2
O at 60 LPM ow
Humidity
Humidity
min
Output: 10 mg H
2
O/L
Measured @ max ow, 35° C, 15% RH.
HeatedTubingSpecications
Maximum Recommended Pressure
25 cm H
2
O
Inner Diameter
15 mm (0.6 in.)
Length
1.83 m (6 ft.)
Heated Tubing Temperature Range
16° to 30º C (60° to 86° F)
Heated Tubing Temperature Cut-out
≤ 41° C (≤ 106° F)
Material
Flexible plastic and electrical components
Electrical (Heated tubing is powered by the attached heated humidier)
Refer to “Electrical” section of DreamStation Heated Humidier Specications
Environmental
Refer to “Environmental” section of DreamStation Heated Humidier Specications