4.13.3 OPP Test Function(Only available for T suffix
version)
This series of electronic loads has an over-power protection (OPP) test
function. In the OPP test mode, when the input voltage reaches the Von value,
the load starts to work after a delay. The current increases by the step value
at regular intervals, and the load input voltage is detected at the same time to
determine whether it is higher than the OPP voltage value. If it is higher, it
means that OPP has not occurred, and the power step operation is repeated
until it reaches the cut-off power; if it is lower, it means that OPP has occurred,
and then check whether the current power value is within the target range. If it
is within the range, it is Pass, otherwise it is Fault.