the VON function is set to a value other than 0. If it is set, please reset the
Von value to the minimum value of 0.
4.12 Compensation Function
When the electronic load operates with high current, significant voltage
drops can occur across the connection cables between the load and the
device under test (DUT). To ensure measurement accuracy, the electronic
load provides a pair of remote compensation terminals. Users can use
these terminals to sample the precise output voltage of the DUT.
Operating Steps:
1. In the system parameter settings page, locate the remote
compensation (RComp) option and select "Rear." Press confirm to
enable the remote compensation function. At this point, the remote
sampling voltage will be used as the actual detected voltage.
2. Return to the parameter settings page and select the remote
compensation (RComp) option as "Front." This will disable the remote
compensation function for the rear terminal, and the load will measure
the voltage at the main front-panel binding posts as the actual voltage.
Note:
After enabling rear remote compensation, ensure correct polarity
connections between the DUT and the terminals. Do not leave
connections floating or reverse the polarity! Do not switch the sense mode
while the system is running!
4.13 Test Function
4.13.1 Battery discharge test function (Only available for
T suffix version)
The electronic load can use the Battery mode to simulate the battery
discharge test. After selecting the discharge test mode, you can set the test
mode termination conditions: "cut-off voltage (Vstop)", "battery capacity
Bekijk gratis de handleiding van Owon OEL15, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.