Je bekijkt pagina 30 van 42
25
Single Pulse Mode (Pulse)
In single pulse mode, when dynamic test operation is enabled, the load
will switch once between the T1 value and T2 value. Each time the load
receives a “SHIF+. trigger command, it will perform a single switching action.
The diagram below shows a schematic of the operation interface in this mode.
4.11 VON Function
When testing some power products with slow voltage rise speed, if you turn
on the input of the electronic load first and then turn on the power supply, the
power supply may be pulled to protect. To this end, the user can set the Von
value. When the power supply voltage is higher than this value, the electronic
load will start to pull the load. For this series of electronic load products, when
the power supply voltage to be tested drops and is less than the Von Point
unloading voltage, the load will not unload.
Note: Please confirm whether you need to set the load voltage. Setting the
load voltage is to facilitate users to limit the working voltage value. If it does
not need to be limited, please do not set it arbitrarily to avoid the trouble of not
being able to load. If the instrument cannot load, please first check whether