24
4.10 Dynamic Mode Function
In dynamic mode, the electronic load allows the user to set two load
current parameters corresponding to the constant mode. By configuring
the operation mode, the load current will switch back and forth between
the two set values.
Dynamic Mode Setting Parameters:
Dynamic Test
Description
Operation Mode
Select the continuous operation mode: Continuous /
Single
T1 Value
Set the T1 load current parameter
T1 Time
Set the T1 load current duration
T2 Value
Set the T2 load current parameter
T2 Time
Set the T2 load current duration
Rise Slope
Set the rise slope
Fall Slope
Set the fall slope
Continuous Mode
In continuous mode, when dynamic test operation is enabled, the load
will continuously switch between the T1 value and T2 value.