Menu Operations
38
The monitor should be provided a stable status before starting the calibration.
That is, before calibration, make sure your monitor has been powered up for a
certain time (we recommend at least 20 minutes), thus, this will warm up the
monitor and make it in an excellent status, and please make sure there is no
intense light bursting on the monitor.
About the Detector Position
Put the detector of your probe at the center of the monitor. as shown in Figure
5.2-22.
You should make sure there is no other light source in your color calibration
system, recommended a totally black environment.
You’d better lean the desktop monitor in a certain degree to be close to the meter.
Figure 5.2-22 Calibration Probe on Screen
◼
Method 1: Autonomous Calibration
Calibration Preparation
:
Connect the calibration probe to the
CALIBRATE
connector of your monitor
using the calibration cable, as shown in Figure 5.2-23.