Menu Operations
39
Figure 5.2-23 Calibration Connector on Mega Monitor
Calibration Steps
First, ensure that the autonomous calibration method is activated. Select
system
→
Advance Setup
→
FACTORY MANAGE
item in the monitor
settings, and confirm this parameter is set to the default state
OFF
. During
autonomous calibration, the system will automatically use the default setting,
and no modification is required for
FACTORY MANAGE
item. Now, you can
start the device's autonomous calibration process.
Next, select
system
→
Advance Setup
→
FACTORY MANAGE
in the monitor
settings. A prompt box will appear, as shown in Figure 5.2-24. Press the
joystick down to select "
OK
" to start autonomous calibration.
Figure 5.2-24 Autonomous Calibration Prompt
Once the calibration starts, select "
START
", as shown in Figure 5.2-25. and
press the joystick down to confirm and proceed with the calibration. A prompt
indicating the probe placement will appear, as shown in Figure 5.2-26.