Testing using the PAT250SX
6.2 To repeat a continuity test (Class 1 and Extension lead tests only or )
Press key during R
CONT
(or R
PE
) test to enable repeat test. The symbol will be displayed.
When the timer symbol has disappeared and the repeat symbol is flashing, press
to run repeat test
Press
to exit repeat test.
6.3 To repeat continuity test with 1.0 Ω limit (NOT available on UK models)
At the end of a FAILED continuity test the symbol will flash for up to 5 seconds.
Press the
or button to repeat the test within the 5 seconds.
The test will be repeated with a 1.0
W pass limit.
6.4 Lock a test in the ON state:
R
CONT
(R
PE
) or R
INS
(R
ISO
) can be locked ON ( ) during a test for up to 3 minutes. To Lock R
CONT
(R
PE
) or R
INS
(R
ISO
) on:
Press
, or during the R
CONT
(R
PE
) or R
INS
(R
ISO
) test
Press key again to unlock test and proceed to next test
NOTE : This feature is available in group test and QT mode.