Go to the “Sound alarm” line then press OK.
Select “YES” to activate the audible alarm or “NO” not to activate it then press OK.
Go to the line Acknowledgement duration” then press OK.
Use Up/Down/Left/Right arrows to enter the value of the alarm acknowledgement in minute then press OK (this
value is between 0 and 60 minutes).
3.5.4 Configure the alarm on Transmitter state mode
The alarm mode “Transmitter state” is selected.
Go to the “Conguration” line then press OK.
Press OK on the “Condition selection” line.
The transmitter displays the list of alarms conditions:
Ambient temperature too high*
Ambient temperature too low*
Memory error
Board error*
Probe 1 error
Probe 2 error
Pressure value too high*
Go to the alarm condition with Up and Down arrows then press OK to select the condition.
The alarm condition is ticked.
Use the right arrow to go to the 2
nd
page of the conditions and use the Left arrow to back to the 1
st
page.
Press Esc key when the required conditions are selected.
The transmitter displays a message asking to save.
Go to YES then press OK.
Then go to the Audible alarm” line then press OK.
Select “YES” to activate the audible alarm or “NO” not to activate it then press OK.
Go to the line Acknowledgement duration” then press OK.
Use Up/Down/Left/Right arrows to enter the value of the alarm acknowledgement in minute then press OK (this
value is between 0 and 60 minutes).
The alarm must be completely set to be active (for example, do not keep “---” in front of a threshold, a delay,...).
3.5.5 Configure the relays (optional)
Press OK.
Enter the security code.
Go to the “Relays” line then press OK.
The transmitter displays the 4 relays: “Relay 1”, “Relay 2”, “Relay 3” or “Relay 4”.
Select the relay to congure then press OK.
Press OK on “Selection” line.
Select “ON”, “OFF” or Alarm 1, 2, 3 or 4” with Up and Down arrows then press OK:
“ON” is selected: the relay is connected. It also allows to perform a test of the relay.
“OFF” is selected: the relay is deactivated.
Alarm 1, 2, 3 or 4” is selected: the activation of the relay is assigned to an alarm previously congured.
Go to “Security” line then press OK.
The outputs relays are by default in negative security: the relay is energized during an alarm condition. Via the
keyboard, it is possible to switch the relays in positive security: the relay is de-energized during an alarm
condition or a power cut.
Select “Positive” or “Negative” with Up and Down arrows then press OK.
18
Congure the transmitter
*Only available with the pressure board