PG17xx Electronic manometer
38 © ifm electronic gmbh 11601190 / 01 07 / 2025
The measuring range is scalable: The measuring range can be limited using the ASP2 and AEP2 parame-
ters.
Minimum distance between ASP2 and AEP2 = 20 % of the final value of the measuring range.
If measured values are outside the display range or in case of an error, messages are displayed (UL, OL,
Err) and the analogue signal goes to a lower or upper limit value (
figure).
The [FOU2] parameter can be used to set the behaviour of the analogue output in the event of an error
Chapter: Error behaviour of the outputs.
Analogue signal with setting ou2 = I:
MEWMAW AEP2
ASP2
[% MEW]
P -1,25 0 100 103,13
4
3,8
20
20,5
21,5FOU2
= ON
FOU2
= OFF
[mA]
ULErr ErrOL
3,5
1
2
3
4
5