OVERVIEW
19
Considerations
The following situations should be taken into consideration when
using the power supply.
Inrush
current
When the power supply switch is first turned on, an
inrush current is generated. Ensure there is enough
power available for the power supply when first
turned on, especially if a number of units are turned
on at the same time.
Pulsed or
Peaked loads
When the load has current peaks or is pulsed, it is
possible for the maximum current to exceed the mean
current value. The GPE-1000 power supply ammeter
only indicates mean current values, which means for
pulsed current loads, the actual current can exceed
the indicated value. For pulsed loads, the current
limit must be increased, or a power supply with a
greater capacity must be chosen. As shown below, a
pulsed load may exceed the current limit and the
indicated current on the power supply ammeter.