Maintenance
11 Maintenance
11.1 Cleaning of Housing
Attention!
!
Life endangering due to electric
shock!
The test adapter is operated with
electric power. Therefore there is a
general risk of electric shock which
can be fatal or cause severe injury.
– The test adapter must be voltage-
free both before and during cleaning.
Switch the test adapter off and dis-
connect it from the mains power
supply.
– The test adapter may not be con-
nected to a measuring circuit before
and during cleaning.
– Never immerse the test adapter in
water or other liquids.
– Never touch the test adapter to
with wet or moist hands.
No special maintenance is required. Keep
outside surfaces clean and dry.
Attention!
!
Unsuitable cleaning agents such as
aggressive or abrasive cleansers re-
sult in damage to the test adapter.
– Use a cloth for cleaning which has
been slightly dampened with water.
– Avoid the use of cleansers, abra-
sives or solvents.
11.2 Technical Safety Inspections
Testing per DGUV Regulation 3
Subject your test adapter to technical safety
inspections at regular intervals.
The test adapter is designed in accordance
with IEC 61010 as a protection category I
and II test instrument.
Testing of the protective conductor, insula-
tion resistance and touch current is
described in the following subsections.
11.2.1 Testing of Protective Conductor Resis-
tance R
PE
Test Adapter Contacting,
Single-Phase (230 V)
Protective conductor resistance is tested
between the PE contact at the mains plug
(inlet plug N1) and PE contacts ICCB IN1
through IN4.
Testing the Single-Phase Connection
➭
Turn the top rotary selector switch
(S1) at
the test adapter to the Status C position
(system state: vehicle ready for
charging).
With the exclusion of the exceptions speci-
fied in the following table, a protective con-
ductor resistance R
PE
of less than 2 is
permissible. This is due to the design of the
test adapter.
Test Adapter Contacting, 3-Phase (400 V)
Protective conductor resistance is tested
between the PE contact at mains connec-
tion N2 and the PE contact at ICCB-IN1
(parallel to IN2, IN3 and IN4).
Step
Rotary
switch
Test
Adapter
ICCB
Protective
Conductor Re-
sistance R PE,
L1
LED
PE
LED
LED Permissible
1 < 2 W
2 < 2 W
3 < 2 W
4 < 2 W
5 < 2 W
6 >30MW
7 >30MW
8 >30MW