Page 45
DWL-5x000XY DATA SHEET
Front Measurement Position
Back Measurement Position
Example sensor position illustration in absolute offset setup for single-axis mode
Example:
Front Measurement Position Value = 1.5 °
Back Measurement Position Value = - 2.3 °
Absolute offset data = (Front Measurment Position Value + Back Measurment
Position Value) / 2
Absolute offset data = (1.5 ° + (- 2.3°)) / 2
= - 0.4 °
Absolute Level mode measurement = Measurement data - Absolute offset data
= 1.5 °- (-0.4 °)
= 1.9°
Note:
Absolute offset data required to perform on predetermined measurement surfaces.
Ensure the device front position measurement & back position measurement are align in the
same position with perfect 180° rotation.
For Dual Axis absolute level offset calculations, the X & Y-Axis require to calculate separately as
the offset value for both axis could be different.
6.5.8. Location Setting*
To set location :
<< (To sensor) : 0x06, 0x01, 0x08, 0x00, 0x00, 0x00, 0x00, 0x5A, 0x00, 0x00, 0x00, 0x00