Brecknell SBI-110 handleiding

29 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 19 van 29
AWT35-501860 Issue AE.1 Page 19
4.2. Supervisor Menu
In the Supervisor Menu there are various submenus available to configure system parameters,
specific sections of the operating modes and the calibration feature (see also chapter 6. Calibration).
4.2.1. Supervisor Menu Levels [LF]
Navigate the options by pressing ZERO (◄), TARE (►) and increase a numeric value by SELECT
(▲) key. Press UNITS (ESC) to exit the option or the menu and PRINT () key to confirm.
4.2.2. Supervisor Functions Descriptions [LF]
There are 8 functions.
LF 1
1
st
Calibration: see chapter 6. Calibration on page 22.
LF 2
Configuration: use this item to configure system parameters and specific sections of the
operating modes. See chapter 5. Scale Configuration on page 21.
LF 3
Linearization: use this to add 1 linearization point for the scale.
CAUTION! Perform the linearity procedure only if test weights applied to the scale
between the zero and span calibration points are showing slight inaccuracies, such as ±
a few divisions. If large inaccuracies are recorded, this indicates a possible mechanical
problem or possible load cell failure which linearity calibration may not be able to correct.
LF 4
ADC Update Rate: use this to select the sampling frequency: 15 (1), 30 (2) or 7.5Hz (3)
Factory default: SPEEd 1 (standard 15Hz)
LF 5
Zero Tracking: use this to define a ± 0 /5 divisions range around zero. When the scale
weight is not at the center of zero but inside this range, ½ of the weight will be subtracted
until that the weight is inside the center of zero region.
Factory default: ZP 0 (off)

Bekijk gratis de handleiding van Brecknell SBI-110, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.

Productinformatie

MerkBrecknell
ModelSBI-110
CategorieNiet gecategoriseerd
TaalNederlands
Grootte6197 MB