AWT35-501860 Issue AE.1 Page 20
LF 6
Factory default: none.
See also Chapter 6 on page 22.
LF 7
Gravitational value: use this item to key in a G constant value: 9.78031 < G < 9.83217
If the scale has been calibrated at a different location and it is not possible to re-calibrate
with known test weights, the scale can be adjusted using this gravity factor.
• Current G value can only be edited if Cal Switch ON. See chapter 6. Calibration.
Factory default: 9.79888
CAUTION! Verify with local agencies if adjusting the gravity factor is accepted in your
area. It may be required that calibration be done with certified weights.
LF 8
Zeroing at Every Power UP: use this item to set an auto-zero at the power-up.
Factory default: SEtZ Y (on)
4.3 Quick Calibration Menu
Before entering this menu, make sure that UF-5 is set to HOLD 0. See chapter 6. Calibration on page
22 for a detailed procedure.
4.3.1. Quick Calibration Menu Levels [ECF]
Navigate the options by pressing ZERO (◄), TARE (►) and increase a numeric value by SELECT
(▲) key. Press UNITS (ESC) to exit the option or the menu and PRINT (↵) key to confirm.
4.3.2 Quick Calibration Functions Descriptions [ECF]
There are 3 Calibration Functions.
ECF-1
Calibration: ZERO / SPAN. Test weight should be at least 1/3 of the max. capacity
ECF-2
Zero Calibration (only)
ECF-3
Span Calibration (only)