AWT35-501860 Issue AE.1 Page 11
2.5.2. Operation Keys
Keys are shown below along with their functions. Some keys have secondary functions.
Press and hold the OFF key to turn off the scale. “oFF” will be displayed for 2 seconds.
Press the ON key to turn on the scale.
The scale will run through a display test before displaying 0.000.
- Press the UNITS key to scroll through the active units of measure.
- In Menu mode: It aborts a numeric entry and acts as an ESCAPE key (ESC).
- Press the ZERO key to zero the scale. Enabled only within the ±2% range of capacity.
- In Menu mode: It acts as a left arrow key (◄).
- Press the TARE key to perform a tare function.
- In Menu mode: It acts as a right arrow key (►).
- Press the SELECT key to toggle between the operating modes enabled.
- In Menu mode: It increments a numeric value (▲).
- Press the PRINT key to send information to a peripheral device.
- In Menu mode: It acts as an ENTER key to accept a displayed value or function ().
2.6. Error Codes
Err H
Initial zero too high (+10% of full cap. see LF-2)
Err L
Initial zero too low (-10% of full cap. see LF-2)
Err N
Unstable internal count, this may indicate the load cell or electronics are faulty
Err Z
Out of Zero range (+2% of full cap.)
Err 4
EEPROM failure, this may indicate the indicator is faulty
-OL-
Overload, full capacity +9d
-UL-
Underload, zero cal. -20d