Commissioning IDS WA System
12 ba76199e03 08/2018
The list includes wireless sensors in the reception area and sensors that are con-
nected to the meter.
The maximum number of sensors (with or without cable) that can be connected to
the meter corresponds to the number of IDS channels of the meter.
Prerequisites for the wireless operation of sensors:
• The IDS meter can administer wireless sensors
(see section 3.2 S
YSTEM REQUIREMENTS)
• The IDS WA-M adapter is connected to the meter
• An IDS WA-S adapter is connected to an IDS sensor
• The battery in the IDS WA-S- adapter is charged
• The radio link is less than 10 m
• The radio link is free of visible obstacles
1 Sensor name, series number (max. 8 sensors)
2 Setting of the connection to the meter (
---/
LED/
)
Open the connection setting for the sensor with
<ENTER>.
The IDS WA-S adapter of the selected sensor briefly lights up red.
3 * (asterisk): The sensor is already connected to another meter. If the
sensor is connected to another meter, the
<ENTER> and <CAL> keys are
blocked for the sensor there. For information, the status indicator [RO] is
displayed in the measured value display.
4[
i
] BAT xx %: Current charging condition of the battery in the IDS WA-S
adapter
5[
i
] SIGNAL xx dBm: Current signal strength
6
<F2>/[ ] or <F4>/[ ]: Update the list of sensors
1
2
4
5
6
3
1 4110W 16010001
LE
D
2 4310W 15080123
---
3 4320W 15120056
---
4* FDO 4410W 15090246
---
[
i
] BAT 86 % SIGNAL -75 dBm
Sensors