YSI 2900D handleiding
Handleiding
Je bekijkt pagina 137 van 172

9-13
Connect Cal bottle tubing marked C2B to the matching fitting on the opposite side of the manifold.
C3A
Apply plenty of grease (included with the new pump tubing) to the longer section of new calibrator pump tubing.
Insert the new pump tubing around the pump roller assembly and into calibrator pump C3A. Make sure the short section
of tubing is on the left side of the pump. The tubing attached this fitting connects to the left fitting on waste module 3.
Place a red pump head cover onto the cal pump and press until it snaps into place.
Connect the tubing from the right side of the pump to the second fitting from the top of the bottom manifold.
Connect Cal bottle tubing marked C3A to the matching fitting on the opposite side of the manifold.
C3B
Apply plenty of grease (included with the new pump tubing) to the shorter section of new calibrator pump tubing.
Insert the new pump tubing around the pump roller assembly and into calibrator pump C3B. Make sure the short section
of tubing is on the left side of the pump. The tubing attached this fitting connects to the right fitting on waste module 3.
Place a red pump head cover onto the cal pump and press until it snaps into place.
Connect the tubing from the right side of the pump to the bottom fitting of the bottom manifold.
Connect Cal bottle tubing marked C3B to the matching fitting on the opposite side of the manifold.
9.4.8 Install Waste Modules
After cleaning, flush the waste modules with copious amounts of authorized rinsing agent (see Section 9.1) to remove any
traces of the disinfecting agent. Install new O-rings in the base plate under each waste module. Reinstall waste module 1
using the three hex screws previously removed. Slide waste modules 2 and 3 over the threaded rods. Connect the ends
of the new calibrator tubing to the fittings on the side of each waste module. Connect CA to the left fitting and CB to the
right fitting of each module.
Calibrator and Waste Tubing (Module 1)
Figure 9-13
9.4.9 Waste Tubing
Slide the new waste tubing onto the large fitting on the side of each waste module.
Connect waste
tubing 1 to fitting
C1A
C1B