9-11
Figure 9-11
Remove the pump tubing from each pump.
Install Buffer Tubing
Apply plenty of buffer grease (included with the new buffer pump tubing) to the new buffer pump tubing. Apply the grease
to the large diameter section of each tubing. Do not apply calibrator pump grease to the buffer pump tubing.
Insert the new pump tubing around the pump roller assembly and into each buffer pump. Make sure the small white fitting
with the small diameter tubing is on the left side of pump 1 (bottom for pumps 2 and 3).
Place a blue pump head cover onto each buffer pump and press until it snaps into place.
For Buffer Pumps 2 and 3, insert the small diameter tubing through the holes in the manifold. Connect the end of the small
diameter tubing to the buffer valves as described below.
Connect Buffer Tubing
Connect the small diameter tubing from Buffer Pump 1 to the inner channel of the bottom valve.
Thread the small diameter tubing from Buffer Pump 2 (Buffer Pump 3 for 2950D-4 only) through the inner channel of the
three tubing holders and connect it to the inner channel of the top valve.
Thread the small diameter tubing from Buffer Pump 3 (Buffer Pump 2 for 2950D-4 only) through the outer channel of the
three tubing holders and connect it to the outer channel of the top valve.
Connect the larger diameter tubing from the top of buffer pumps 2 and 3 to the top fitting of each manifold.
Connect Buffer bottle tubing marked B2 and B3 to the matching fittings on the opposite side of the manifolds.
9.4.7.3 Calibrator Pump Tubing
C1A
Note that the tubing for C1A is slightly longer than C1B.
Apply plenty of calibrator grease (included with the new calibrator pump tubing) to the longer section of new calibrator
pump tubing. Do not apply this calibrator pump grease to the buffer pump tubing.
Insert the new pump tubing around the pump roller assembly and into calibrator pump C1A. Make sure the short section
of tubing is on the left side of the pump. The tubing attached to this fitting connects to the left fitting on waste module 1.
Place a red pump head cover onto the cal pump and press until it snaps into place.
C1B
Apply plenty of grease (included with the new pump tubing) to the shorter section of new calibrator pump tubing.
Insert the new pump tubing around the pump roller assembly and into calibrator pump C1B. Make sure the short section
of tubing is on the left side of the pump. The tubing attached this fitting connects to the right fitting on waste module 1.
Place a red pump head cover onto the cal pump and press until it snaps into place.

Bekijk gratis de handleiding van YSI 2900D, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.

Productinformatie

MerkYSI
Model2900D
CategorieNiet gecategoriseerd
TaalNederlands
Grootte50780 MB