4.6 Short Circuit Simulation Function
The electronic load can simulate a short circuit at the input terminal to test
whether the protection function of the device under test (DUT) operates
correctly when a short circuit occurs at its output terminal.
To switch to the short circuit state, press (SHIFT) + 1 (SHORT) on the front
panel of the electronic load. The short circuit operation does not affect the
current settings, and pressing (SHIFT) + 1 (SHORT) or MODE again will
return the electronic load to its original set state. The actual current consumed
by the electronic load during a short circuit depends on the current operating
mode and current range. In CC , CP, and CR modes, when the current is
within the maximum range, the short-circuit current will be equal to the current
input value. If the current exceeds the maximum range, the maximum short-
circuit current will be 110% of the current range. In CP mode, a short circuit is
equivalent to setting the loads constant voltage value to 0V.
Figure 4.6-1 Short circuit simulation interface