2
Dynamic Mode Setting Parameters:
Dynamic Test
Description
Operation
Mode
Select the continuous operation mode: Continuous /
Single
T1 Value
Set the T1 load current parameter
T1 Time
Set the T1 load current duration
T2 Value
Set the T2 load current parameter
T2 Time
Set the T2 load current duration
Rise Slope
Set the rise slope
Fall Slope
Set the fall slope
Continuous Mode
In continuous mode, when dynamic test operation is enabled, the load will
continuously switch between the T1 value and T2 value.
Single Pulse Mode (Pulse)
In single pulse mode, when dynamic test operation is enabled, the load