Use longer record lengths to
either capture more samples
for a waveform record, or
capture more cycles of a
waveform, then use the Zoom
controls to investigate the
waveform in more detail.
Maximum zoomed waveform
record captured with 5000
points
Maximum zoomed waveform record captured with 5000 points
Using the roll display mode
Roll mode gives a display similar to a strip chart recorder
, in that the waveform moves slowly, or rolls, from right to left on the screen.
Roll mode is used to display low-frequency signals. Roll mode lets you see acquired data points without waiting for the acquisition of a
complete waveform record.
Roll mode is not a button- or menu-selected mode, but is present when the trigger mode is Auto and the horizontal scale and record
lengths are set as follows:
Table 5: Roll mode is enabled when:
Horizontal scale setting Record length (number of samples)
40 ms/div 1000, 2000, 20k
Roll mode tips
Push Run/Stop to halt Roll mode.
Roll mode is disabled under the following conditions:
When using Reference, Math, or FFT waveforms
Using the Zoom function
Taking measurements (Measure menu)
Acquisition setup
TBS1000C Series Oscilloscopes User Manual 74