TROUBLESHOOTING
Problem
Possible cause
Solution
Droplets left inside
the tip leakage or
pipetted volume too
small
Unsuitable tip
Non-uniform wetting of the
plastic
Tip incorrectly attached
Foreign particles between tip
and tip cone
Instrument contaminated
Insufficient amount of grease
on piston and O-ring
O-ring not correctly
Positioned or damaged
Incorrect operation
Calibration altered or
unsuitable for the liquid
Instrument damage
Use original tips
Attach new tip
Attach firmly
Use original tips
Clean the tip cone, attach
new tip
Clean and grease O-ring and
piston, clean the tip cone
Grease accordingly
Change the O-ring
Follow instructions carefully
Recalibrate according to
instructions
Send for service
Push button jammed
or moves erratically
Piston contaminated
Penetration of
solvent vapors
Clean and grease O-ring and
piston, clean the tip cone
Pipette blocked;
aspirated volume
too small
Liquid has penetrated t ip
cone and dried
Clean and grease O-ring and
piston, clean the tip cone
Tip ejector jammed
or moves erratically
Tip cine and/or ejector collar
contaminated
Clean the tip cone and the
ejector collar