1. Hold down the tip pipette.
2. Place the tooth of the opening
tool between the tip ejector and
the tip ejector collar to release
the locking mechanism. (D)
3. Carefully release the tip ejector
and remove the ejector collar.
4. Place the wrench end of the
opening tool over the tip cone,
turning it anticlockwise. Do not
use any other tools. (E)
5. Wipe the piston, the O-ring and
the tip cone with ethanol and a
lint-free cloth
Note: Models up to 10ul have a fixed
O-ring locked inside the tip cone.
Therefore, the O-ring cannot be
accessed for maintenance.
6. Before rep lacing tip cone it is
recommended to grease the
piston slightly using the silicone
grease provided
Note: Excessive use of grease may
jam the piston.
7. After reassembling use the
pipette (without liquid) several
times to make
8. Check the pipette calibration
(D)
(E)