Basics 11
Stanford Research Systems DC215 Voltage/Current Source
S Trig: Indicates that a step trigger has occurred.
Meas: Indicates that a measurement is active
S Rate: A warning that indicates a step trigger has occurred while the on-going
step is still in process.
M Rate: A warning that indicates a measurement trigger has occurred while an
on-going measurement is still in process.
Keypad Area
The keypad area contains cursor keys, numeric keys, control keys and indicators.
Cursor Keys
The cursor keys are used to adjust setting values or to navigate within the menu
system.
See Display and Entry Behavior (page 14) for details on using the cursor keys.
Numeric Keys
The numeric keys include 0 9, ±, decimal point and [CLR] (clear). There are two Enter
keys: [mV/mA] and [V/A] that are used for both united and non-united entry.
See Display and Entry Behavior (page 14) for details on using the numeric keys.
The numeric keys in conjunction with [SHIFT] are used to access secondary functions
and the menuing system.
Control Keys
The control keys perform various actions as well as bringing different parameters to the
entry screen for modification. There are indicators associated with some of the control
keys.
[ON/OFF]
The [ON/OFF] key is one of the controls that can switch the output On or Off. See
ON/OFF Behavior (page 22) for more details. The On indicator is lit when the output is
on regardless of the method used to turn it on.
Note: [ON/OFF] can be overridden by the Output Enable. See Output Control (page 23)
for details.
[V/I]
This toggles between the voltage and current source type (voltage or current).
When changed, a source type message is briefly displayed in the left display, the units
indicators and decimal point are updated and the output is turned off.
Note: After toggling between voltage and current, the output range is set to the last
source value and range for that source type, indicated by the units and decimal point.