Guidance and manufacturer’s declaration – electromagnetic immunity
The AC-120MD is intended for use in the electromagnetic environment specified below. The customer or
the
user of the AC-120MD should assure that it is used in such an environment.
Immunity test
IEC 60601
te
st level
Compliance
level
Electromagnetic environment – guidance
Portable and mobile RF communications
eq
uipment should be used no closer to any part
of the AC-120MD, including cables, than the
recommended separation distance calculated
from the equation appliance to the frequency of
the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
outside ISM
ba
nds
c
3 Vrms
d
= 1.2
P
6 Vrms
150 kHz to
80 MHz
in ISM bands
c
6 Vrms
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.
7 GHz
3 V/m IEC 60601-1-2: 2007
d
= 1.2
P
80 MHz to 800 MHz
d
= 2.3
P
800 MHz to 2.5 GHz
IEC 60601-1-2: 2014 + A1: 2020
d
= 2.0
P
80 MHz to 2.7 GHz
Where
P
is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
d
etermined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the
vicinity of
equipment marked with following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
a
bsorption and reflection from structures, objects and people.