Guidance and manufacturer’s declaration – electromagnetic immunity
The AC-120MD is intended for use in the electromagne
tic environment specified below. The customer or
the user of the AC-120MD should assure that it is used in such an environment.
Immunity test
IEC 60601
te
st level
Compliance level Electromagnetic environment – guidance
Electrostatic
di
scharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, a
relative humidity of at least 30% is
recommended.
Electrical fast
tran
sient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
o
utput lines
±2 kV for power
s
upply lines
±1 kV for input/
o
utput lines
Mains power quality should be that of a typical
co
mmercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line
(s)
±2 kV line(s) to
ea
rth
±1 kV differential
mod
e
±2 kV common
mod
e
Mains power quality should be that of a typical
co
mmercial or hospital environment.
Voltage dips,
short
inte
rruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
0%
U
T
(100% dip in
U
T
)
for 0.5/1 cycles
a
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25/30 cycles
a
(for 0.5 sec)
0%
U
T
(100% dip in
U
T
)
for 250/300
cycle
s
a
(for 5 sec)
0%
U
T
(100% dip in
U
T
)
for 0.5/1 cycles
a
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25/30 cycles
a
(for 0.5 sec)
0%
U
T
(100% dip in
U
T
)
for 250/300
cyc
les
a
(for 5 sec)
Mains power quality should be that of a typical
co
mmercial or hospital environment. If the user
of the AC-120MD requires continued operation
during power mains interruptions, it is
recommended that the AC-120MD be powered
from an uninterruptible power supply or a
battery.
Power frequency
(50
/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Proximity
ma
gnetic fields
IEC 61000-4-39
65 A/m
134.2 kHz
Pulse modulation
2.
1 kHz
7.5 A/m
13.56 MHz
Pulse modulation
50 kHz
65 A/m
134.2 kHz
Pulse modulation
2.
1 kHz
7.5 A/m
13.56 MHz
Pulse modulation
50 kHz
Proximity magnetic fields of the AC-120MD
should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
a For example, 10/12 means 10 cycles at 50 Hz or 12 cycles at 60 Hz.