XPT800 Sound Level Meter and Spectrum Analyzer
V1.0
Page 75 of 124
Log file max size (Log_file_size)
Limits the size of files generated during continuous recording to a user-selectable value in MB (1, 2, 5,
10, 20, 50, 100MB). Files thus generated within the folder are named, for example, as histo-
ry_0001.dodl, history_0002.dodl. An automatic queuing function present in the management soft-
ware and in the NS-Storage portal allows the records divided as follows, to be merged into a single
trace.
Log synchronization (Log_sync)
Activates the automatic data synchronization mode on cloud. The function can be activated directly
with the button in the drop-down menu. If log_sync is on, storage support cannot be accessed via
USB
Sync files (Log_sync_type)
Allows selection of file types to be synchronized to cloud among: Reports, Events, Globals, Audio.
Start of Reports (Report_sync)
If set to "CLOCK" it aligns the calculation time window of Report parameters with the nearest
"round" time. For example, if Report_Step set to 1h and Report_sync set to clock, the integration time
window will go from 9:00 to 10:00, 10:00 to 11:00, and so on. If Report_sync set to "MANUAL" and
Report_step set to 1h, the integration window will go from the time of measurement start (e.g.
08:45:00) to the next hour (e.g. 09:45) and so on.
Sync backlog (Sync_files)
View the list of files waiting for synchronization
10.6.2 Continuous datalogging of Time History group
Continuous recording of the "time history" group includes subgroups of Instantaneous (Inst), Maxi-
mum and Minimum (Mx Mn), Integrated (Avg) and Auxiliary (Aux) parameters.
For each subgroup it is possible to select by checkbox one or more parameters that are stored with
user-defined step (menu SETUP > History interval), displayed at the top of the parameter selection
panel.
When for the "Time history" group, the user selects at least one parameter, the log function of the in-
strument, stores in the memory at least one file of the type history.dodl. A folder identified by date
and time, containing *.dodl files, is created with each measurement. The folder may contain other
types of files with specific functions
Info
For more details on the acoustic parameters (Parameter Classes) that can be stored
through the "time history" group, please refer to the section 19 Appendix B Parameters
on page 104.