PRO 915-2 Differential manometer / Thermometer
Page 21 of 30
7.5 General warnings on probes usage
Wait for thermal equilibrium between the sensitive part of the probe and the area to be meas-
ured before taking the measurement.
Thoroughly clean the probe after use.
Danger!
If the probe has a metal stem or other metal parts, be careful not to come into contact with live
parts.
Caution!
Do not expose the probe to corrosive gases or liquids!
Do not expose the probe to temperatures exceeding the operating limit specified for the
probe, the measuring sensor may be damaged.
Avoid performing measurements in the presence of high-frequency sources, microwaves
or strong magnetic fields.
Connect the probe to the instrument without forcing or bending the connectors or
contacts!
Do not deform or drop the probe!
Bending within moderate radius is only allowed for mineral insulated probes.
7.6 Warning on USB port isolation
The instrument USB port is not galvanically isolated, and its connection, either to the PC or to
an external power supply that is not isolated from ground, may affect the measurement when
using a probe that is in contact with the measured medium. Under such conditions, always
check the measurement with and without USB connected: if a difference in the detected value
is noticed, disconnect USB, or use an external suitable USB isolator (for example, DFRobot
FIT0860) to detect reliable measurements.