ECO 410 / ECO 410T - O2-Analyser / oxygen meter
Operation
B-H86.0.31.DK2-2.1 Page 11 of 22
Parameter
Values
Meaning
ON
No automatic shut off of the backlight
IN,T
Factory settings
NO
Use current configuration
YES
Reset device to factory settings.
After confirming with the function-button, the
display shows: IN,T DONE
Extended notes on the setting parameters:
1 Operational depth: max. depth in meters at which, with the mixture used
(e.g. Nitrox 32), the set maximum oxygen partial pressure "Po2"
(usually 1.4 bar) is reached.
2 Air pressure: is necessary for correct calibration and calculation.--
3 Max. Partial pressure: Due to the toxicity of oxygen, it is recommended not to be
exposed to an oxygen partial pressure above 1.4 bar. This applies
under optimal conditions and health. De-pending on the
condition of the day and stress (e.g. cold water), this limit should
be reduced further down (e.g. 1.2 bar)!
4.2 Changing the sensor
1 sensor
2 sensor cable
3 protective sensor cover
Pull the sensor with the connected sensor cable out of the protective sensor cover.
Disconnect the used sensor and dispose it properly.
Open the tin can with the new sensor.
Remove the sensor from the tin can.
Connect the sensor to the plug of the sensor cable.
Then push the connected sensor into the protective sensor cover.