ECO 410 / ECO 410T - O2-Analyser / oxygen meter
Operation
B-H86.0.31.DK2-2.1 Page 13 of 22
4.4 Measurement on immersion gas cylinders
Note
The combination adapter for compressed air and NITROX (GZ-5826) as well as the T-piece
are not included in the standard accessories of every instrument version, but can be
purchased separately (see chapter 10.2 Accessories on page 22).
The instrument should be calibrated first, after that a concentration of about 20.9 % must
be displayed.
1 Screw the combination adapter (GZ-5826)
onto the immersion valve of the immersion
gas cylinder and tighten hand-tight.
2 Screw the flow diverter onto the sensor
and hand-tighten (make sure that there is
an O-ring on both the sensor and the flow
diverter).
3 Plug the T-piece (middle connection) onto
the flow diverter.
4 Plug the T-piece (right or left connection)
onto the combination adapter.
5 If no combination adapter is available, hold
the T-piece directly on the diving valve of
the diving gas cylinder.
6 Slightly open the valve of the diving gas
cylinder until a slight outflow noise can be
heard.
7 Wait until a stable measured value is
displayed. This value is then the desired
measured value of the oxygen
concentration or the MOD value (only
adjustable with the ECO 410T variant).