ECO 410 / ECO 410T - O2-Analyser / oxygen meter
Operation
B-H86.0.31.DK2-2.1 Page 12 of 22
Caution! Risk of injury!
The lid of the opened tin can is sharp-edged!
Note
The sensor, in particular the sensor membrane, must be protected from dirt and moisture.
If the sensor gets wet or falls into water, the sensor should be pulled out of the protective
sensor cover to dry. Measurement can only be resumed after drying.
Disconnecting the T-piece and pulling the flow diverter makes it easier to grip the sensor.
4.3 Adjustment to the sensor signal
Note
Before calibration, the current air pressure should be entered in the device menu. The entry
is made in the PAbS parameter and is entered in hPa without decimal place.
For safe measurements, the function of the sensor and device must be checked regularly:
Regular calibration on ambient air (20.9 vol. % O2) in accordance with the instructions
In the case of higher oxygen values, check the function with a known gas concentration,
preferably in the area of the gas to be measured (e.g. pure oxygen, NITROX50 or similar,
"bump test")
The device displays a sensor evaluation after calibration.
This can give conclusions about the durability of the sensor.
If the evaluation is below 30 %, for example, the sensor should be replaced.
Carrying out the calibration:
1 Press the Function button for approx. 5 seconds until CAL appears in the display to call up
the Sensor Calibration.
2 Release the button, CAL Air appears and after completion for approx. 5 seconds the
sensor evaluation in 10% steps.